Comparison of Custom Cranial Implant Source Data: Manual, Mirrored and CAD generated skull surfaces

Winder, John, McKnight, Wesley, Golz, Tobias, Giese, Alf and Busch, Luder (2006) Comparison of Custom Cranial Implant Source Data: Manual, Mirrored and CAD generated skull surfaces. In: Medical Image Computing & Computer Assisted Intervention, Copenhagen. MICCAI. 98 pp. [Conference contribution]

Full text not available from this repository.

Abstract

One of the objectives of performing cranioplasty is the creation of a satisfactory cosmetic outcome. The purpose of this work is to compare the accuracy of skull defect contours generated by CT volume mirroring and CAD. 3D multi-slice CT scanning was carried out on a cadaver before and after the creation of a large lateral defect. Two volume mirroring and two CAD methods were used to create customised repairs. 3D difference maps were used to compare the repair contour with the original skull. The CAD contour deviation ranged from 0.0 mm to 2.0 mm with 80% of the total defect area less than 0.66 mm as measured by difference maps whilst mirror imaging produced deviations of up to 3.0 mm from the original surface. CAD techniques can be used to produce contours for the repair of cranial defects with minimum deviation from the original skull contour.

Item Type:Conference contribution (Paper)
Faculties and Schools:Faculty of Computing & Engineering
Faculty of Computing & Engineering > School of Engineering
Faculty of Life and Health Sciences
Faculty of Life and Health Sciences > School of Health Sciences
Research Institutes and Groups:Institute of Nursing and Health Research > Centre for Health and Rehabilitation Technologies
Institute of Nursing and Health Research